252. The Harmony 600 has the same poll and does not have the latch

Section 250 left "whether arch 14 latches the same way" open, on the ground that its display routine and its validator have the same shape and neither its poll nor its flag had been read. Both are read now, and the resting state of a connected Harmony 600 is measured. The answer is that the poll, the flag and the validator all transfer instruction for instruction, and the one condition the latch rests on does not.

The same three things, at the Harmony 600 build's addresses

The table below is the 600 image's and this section first gave it as arch 14's, which section 257 corrects: the Harmony 700 carries the same structure with not one shared address, and that section tabulates both. What transfers between the two builds is the verdict bit, the container select bit and the discriminator's address. The finding is stronger for it, since the claim now rests on two images.

arch 12 (Harmony One) arch 14 (Harmony 600)
flags byte 0x1A4 0x68B
verdict bit 2 2
container select bit 3 4
container validator 0x28D92 0x151C6
checksum disagrees, agrees 0x28FFA, 0x29006 0x153DE, 0x153EA
screen number 0x22F, 0x230 0x0B5, 0x0B6
discriminator 0xD04 0xD04
the polled re-check 0x2906C 0x1544A
armed flag 0x318 0x743

The validator's tail is the same code twice. The checksum's two arms clear or set the verdict bit and write 1 to the discriminator on the failing side; then the screen block loads 26 or zero into the screen number and calls the shower. Section 249 read that on both images already.

What section 250 did not notice is that the container select bit is also the screen guard. On both architectures the instruction immediately after the checksum arms is a BTFSS of that bit which jumps past both screen calls when it is clear, 0x2900A on arch 12 and 0x153F0 on arch 14. So a validation of the container the bit does not select shows nothing whatever it finds, which is why validating the safe mode container at boot puts no screen up. That is a second guard in front of the screens, and section 250 described only the verdict one.

Measured on the connected Harmony 600

Read over USB on 4 September 2026 from a Harmony 600 running its application, firmware 0.2, skin 71, software type 0:

address value reading
0x68B 0x16 verdict bit set, container select bit set
0x6E3 0x01 the gate variable, on its arming side
0x743 0x01 the flag is armed

So a connected Harmony 600 rests exactly where a connected Harmony One rests, section 251: verdict standing, flag armed. And its verdict is standing, which says the validator has run over its configuration, so whatever section 110 measured about the journal's five variables being zero on a connected 600, validating is not the same act as loading.

The difference, and it is the one that matters

Arch 12's arming half tests the verdict:

2906c: MOVF PORTA,W; ANDLW 0x10; BNZ 0x29082    the cable
29072: MOVF 0x1a4,W; ANDLW 0x04; BZ  0x29082    the verdict
2907a: MOVLW 0x01; MOVWF 0x318                  arm

Arch 14's does not:

1544a: MOVF 0x6e3,W;      BZ 0x15460            the gate variable
15450: CALL 0x19486; IORLW 0x00; BZ 0x15460     a routine
15458: MOVLW 0x01; MOVWF 0x743                  arm

0x19486 clears a 24 bit index, and for each of 32 steps of two reads a 16 bit word from 0x01F4 plus the index through 0x19C7C, complements both halves, and returns one as soon as it finds a word that is not 0xFFFF, or zero when all 32 are. It asks whether a configuration is present, not whether one was ever believed, and it does not read 0x68B at any depth.

So the latch does not transfer. Section 250's argument is that a failed validation leaves the verdict clear and the flag cleared, and that arming needs the verdict standing, so the arming condition can never be met again and only a power cycle validates. On arch 14 the arming condition does not mention the verdict, so a failed validation re-arms on the next pass and the re-check runs again. A Harmony 600 should therefore not have the stuck screen a Harmony One has, and the battery pull three sections recorded is an arch 12 fact rather than a Harmony fact.

What the gate variable is, deliberately not claimed

Arch 12 reads PORTA bit 4 in the poll itself. Arch 14 reads a variable, 0x6E3, which has one MOVWF writer and two CLRF writers: it is set to 1 at 0x156AC when a state variable holds 4, and cleared at 0x1563C and at 0x154A6, the second being inside the routine the re-check itself calls after both validations, next to a write of UCFG.

That is consistent with it standing for a live USB session, which would give the arch 14 poll the same sense as arch 12's corrected one, arm while connected and re-check while not. It is not asserted: the reading is a comparison with one other architecture plus a state number, and the resting value of 1 on a connected remote is a measurement of where the remote sits rather than of what the variable means.

Why the arch 14 experiment cannot be run

Section 251's runs turned on sending WRITE_MISC selector 0x02 to clear the verdict and watching it come back. That is a write command, and arch 14 has no write target on this bench, so nothing here may clear a Harmony 600's verdict. (Since section 281 arch 14 has one, the Harmony 650, and the 600 still may not be written, so the sentence's conclusion stands for the 600.) The structural reading above stands as a reading, and the only hardware evidence for it is the resting state.

Where it lands

  • tests/test_status_screens.py: the arch 14 poll's three gates and its flag, the container select bit being the screen guard on both architectures, and the negative that arch 14's arming half does not read the verdict bit, with arch 12's arming half as the positive control.
  • Section 250's open item, answered.

Section 251 | the archive | section 253. Section 252 of the findings archive, which is frozen; the reference holds the current answer. The archive on GitHub.